Biacore T system. Biacore™ T from GE Healthcare is a high performance system for real-time biomolecular interaction analysis, using surface . The BIAcore T instrument is located in Swann Rm. Please follow the link for information regarding the Booking and rules for use of the BIAcore T Introduction. This Getting Started handbook is designed as a self-study guide to introduce you to the basic operations of BiacoreTM T, Biacore T Control.

Author: Karg Zulkile
Country: Netherlands
Language: English (Spanish)
Genre: Love
Published (Last): 18 March 2004
Pages: 160
PDF File Size: 12.14 Mb
ePub File Size: 15.6 Mb
ISBN: 457-3-17105-276-2
Downloads: 73754
Price: Free* [*Free Regsitration Required]
Uploader: Samujas

Biaxore follow the link for information regarding the Booking and rules for use of the BIAcore T instrument: Download technical information on the T system. The underlying physics of the principles of SPR are complicated.

However, a working knowledge of the technique doesn’t require a detailed understanding of the theory. To allow the detection of an interaction one molecule termed by BIAcore as the ligand is immobilized onto the sensor surface. Its binding partner termed by BIAcore as the analyte is then injected in aqueous solution ideally with the same components and composition as the running buffer through the flow cell, under continuous flow.

As the analyte binds to the ligand the accumulation of protein on the sensor surface causes an increase in refractive index.

Biacore T – GE Healthcare Life Sciences

This refractive biacoree change is measured in real time sampling in a kinetic analysis experiment is taken every 0. This background response must be subtracted from the sensorgram to obtain the actual binding response. One RU biacord the binding of 1 pg of protein per square mm.

  IEEE STD 802.3AP-2007 PDF

To get round this technical difficulty, BIAcore sensor chips have a – nm thick carboxymethylated dextran matrix attached. However, having very high levels of ligand immobilised on the surface has two important implications for the technique.

Surface Plasmon Resonance (SPR) -BIAcore T

Such high ligand density may result in the rate at which the surface binds the analyte exceeding the rate at which the analyte can be delivered to the surface this latter consideration is referred to as mass transport. If this condition arises, mass transport becomes the rate-limiting step, and as a consequence, the measured apparent on-rate will be slower than the true on-rate.

A related problem is that, following dissociation of the analyte, it can rebind to the unoccupied ligand before diffusing out of the matrix and being washed from the flow cell.

Consequently, the measured apparent off-rate is slower than the actual off-rate. The SPR optical unit consists of a source for a light beam that passes through a prism and strikes the surface of a flow cell at an angle, such that the beam is totally reflected. Under these conditions, an electromagnetic component of the beam, the evanescent wave, propagates into the aqueous layer and can interact with mobile electrons in the gold film at the surface of the glass.

At a particular wavelength and incident angle, a surface plasmon wave of excited electrons the plasmon resonance is produced at the gold layer and is detected as a reduced intensity of the reflected light beam. In Biacore instruments, monochromatic light in the shape of a wedge a broad distribution of incident angles is used.

Each angle of the reflected beam strikes the instrument’s detector at a different point and, thus, the detector continuously records the position of reduced light intensity and calculates the SPR angle t20 that figure.

  FLUKE 289 FVF PDF

Biacore T200

The optical device has no moving parts and the fixed geometry enhances stability and allows binding to be monitored in real time. The SPR angle is sensitive to the composition of the layer at the surface of the gold.

A baseline SPR angle is first determined by washing buffer over the surface with a fixed amount of ligand attached. To this flow of buffer, some analyte is then added.

The binding of analyte to immobilized ligand causes an increase in the refractive index at the surface, thereby changing the SPR angle because it is directly proportional to the amount of bound analyte. The exact relation between RU and ng of material bound will vary with the refractive index of the analyte.

If the added molecule does not bind to a target or receptor, the SPR angle change in the sample and reference flow cells will be the same, and, after subtraction, will give a zero net RU response that indicates no binding occurred. Only bound protein generates a positive SPR signal and that signal, recorded over time, produces a sensorgram. Surface generation and results: The University of Edinburgh is a charitable body, registered in Scotland, with registration number SC